User Manual
2.3 Attaching the accessories
766 IC Sample Processor
13
2.3.4 Installing the splash protection
To avoid any danger of injury by the needle, the 6.2751.040 splash
protection must always be installed when operating the 766 IC Sam-
ple Processor!
1
Remove holding screws
Remove the holding screws 2929 and the washer mounted on the
screw threads 1111 at tower 1919 using the 6.2621.100 Allen key.
2
Remove protective film from splash protection
Remove the plastic film glued on both sides of the splash pro-
tection 11.
3
Install splash protection
Attach splash protection 11 onto the screw threads 1111 at the
tower 1919 and fix it with the holding screws 2929 and the washer
using the 6.2621.100 Allen key.
2929
2929
11
Fig. 5: Installing the splash protection
2.3.5 Installing the needle
1
Remove PEEK compression fitting 5
Remove PEEK compression fitting 55 screwed onto the needle
holder 44.
2
Insert needle
Insert needle 22 (6.2624.000 steel needle or 6.1835.000 PEEK
needle) completely into the opening of the needle holder 44.
3
Fix needle
Tighten compression fitting 55 in needle holder 44 by hand (never
use tools!).
11 Splash protection
(6.2751.040)
2929 Holding screws