User Manual

6 Appendix
766 IC Sample Processor
118
Required Instruments:
Only required, if the RS232 or remote interface is to be checked:
3.496.8550 Test plug (on the "Remote" plug)
3.496.8480 Test plug (on the "RS 232" plug)
6.4.2 Preparing the instrument
Disconnect power plug.
Remove cables to the "RS232" and "Remote" interfaces.
Plug in power plug and immediately press <9> and keep it de-
pressed until the start test pattern disappears.
Alternatively, the <RESET> key can be pressed to cause a reset
followed by pressing (within 0.4 sec) and holding down the <9> key
to open the diagnosis menu.
Switch on and
9
LIFT
or
RESET
CLEAR
and
9
LIFT
Main menu Diagnosis:
diagnosis
>RAM initialization
Open the submenu with <ENTER>
diagnosis
>RAM test
Use <
é
> or <
ê
> to access one menu
item up or down
diagnosis
>display test
Use <HOME> or <END> to access first
or last menu item
diagnosis
>display contrast test
<QUIT> returns to the normal state
diagnosis
>key test
diagnosis
>remote test
diagnosis
>RS232 test
diagnosis
>external bus test
diagnosis
>beeper test
diagnosis
>rack code test
diagnosis
>function test
diagnosis
>instrument number
diagnosis
>power on reset
The "function test" is a Metrohm internal test which should be run only
by service personal. Before starting this test, the needle must always
be removed from the needle holder, otherwise it would be damaged.