User Manual

5 Interfaces
766 IC Sample Processor
100
Object Meaning Entry range/Selection
³ ³
³ Ã.R1O1Off Offset in 1/10-ang. degrees for positions in row 1 03599
³ Ã.R2O2Off Offset in 1/10-ang. degrees for positions in row 2 03599
³ Ã.Num Number of positions (n) 1200
³ Ã.1 Position 1
³ ³ À.Value
³ :
Position in 1/10-angular degrees 03599
³ À.n Position n
³ À.Value Position in 1/10-angular degrees 03599
³
À.RSset
³
RS232 interface
$G effects all the RS settings.
$G
Ã.Baud
³
Baud rate in bit/s 300, 600, 1200, 2400, 4800,
9600
Ã.DataBit Number of data bits 7, 8
Ã.StopBit Number of stop bits 1, 2
Ã.Parity Parity none, odd, even
³.Handsh
³
Handshake
HWs, HWf, SWchar, SWline,
none
À.CharSet Character set IBM, Epson, Seiko, Citizen,
HP
&Info
&Info Instrument information
³
Ã.Report Send formatted reports $G
³ À.Select
³
Selection of the report config, param, usermeth, all
À.ActualInfo Current information
Ã.Lift Lift station
³ À.1 Lift 1
³ Ã.Exist Availability read only
³ Ã.MaxHeight Maximum stroke path read only
³ À.ActHeight Current lift position read only
³
Ã.Rack Sample rack
³ Ã.Code Rack code read only
³ Ã.Type Rack type read only
³ Ã.WorkHeight Working position read only
³ Ã.RinseHeight Rinsing position read only
³ Ã.ShiftHeight Shift position read only
³ Ã.SppecialHeight Special position read only
³ À.ActPos Current beaker position read only
³
Ã.Pump Pump
³ À.1 Pump 1
³ À.State Current pump state (on, off) read only
³
Ã.Inputs Remote inputs
³ À.Status
³
Status of the 8 input lines (Input 07; 1 = on, low,
active; 0 = off, high, inactive) in byte form
read only
³
³
³
³
Status
n
n
=
=
2
0
7