User Manual
5.2 RS232 interface
766 IC Sample Processor
99
Object Meaning Entry range/Selection
³
Ã.FinalSeq Final sequence –
³ Ã.1 Line number of the command –
³ ³ Ã.Cmd
³ ³ ³
³ ³ ³
³ ³ ³
Command (see Start sequence) NOP, MOVE, LIFT, SAMPLE,
PUMP, SCAN, CTRL, WAIT,
ENDSEQ
³ ³ À.
*
³
*
³ :
³ ³
³ ³
Parameter for command (see Start sequence) .MOVE…, .LIFT…,
.SAMPLE…, .PUMP…,
.SCAN…, .CTRL…, .WAIT…,
.END…
³ À.100 Sequence end NOP
³
Ã.Changer Changer settings –
³ Ã.RackNo Rack number 0…16
³ Ã.L1Rate Lift speed 3…1212 mm/s
³ À.ShRate Turning speed of the rack 3…2020 ang. degrees/s
³
À.ManStop Reaction to manual stop –
Ã.RemCtl Command via remote interface 14 bit (1, 0 or *)
À.RSCtl Command via RS232 interface 14 ASCII characters
&Config
&Config Instrument configuration –
³
Ã.Aux General instrument settingsGeneral instrument settings –
³ Ã.Language
³ ³
Dialog language english, deutsch, francais,
español
³ Ã.Contrast Display contrast 0…3…7
³ Ã.Beeper Beeper on/off on, off
³ Ã.DevName Device label 8 ASCII characters
³ Ã.Prog Number of program version read only
³ À.MaxLift Maximum lift height 0…125 mm
³
Ã.RackDef Rack definitions –
³ Ã.RackNo Rack number 1…16
³ Ã.Code Rack code 000001…111111 (6 bit)
³ Ã.Type Rack type M129-2, …
³ Ã.WorkH Working position 0…125 mm
³ Ã.RinseH Rinsing position 0…125 mm
³ Ã.ShiftH Shift position 0 mm
³ Ã.SpecialH Special position 0…125 mm
³ ³
³ À.SpezBeak Special beaker positions –
³ Ã.1 Special beaker 1 –
³ ³ À.Pos
³ :
Beaker position
0…number of rack positions
³ À.8 Special beaker 8 –
³ À.Pos Beaker position 0…number of rack positions
³
Ã.PosTab Position table –
³ Ã.Idx
³
*
Index of the table.
Each table contains the following data:
0…31
³ Ã.Name Name position table 8 ASCII characters
³ Ã.R1Num Highest beaker position in row 1 2…(R2Num – 2)
³ Ã.R22Num Highest beaker position in row 2 (R1Num + 2)…(R3Num – 2)
³ Ã.R33Num Highest beaker position in row 3 (R12Num + 2)…200