User Manual
5.2 RS232 interface
766 IC Sample Processor
95
5.2.3 Trigger
Triggers initiate an action at the 766 IC Sample Processor, e.g. se-
quence start or data transmission. Triggers are marked by the intro-
ducer '$'.
The following triggers are possible:
$G Go Starts processes, e.g. starting the mode run or set-
ting the RS232 interface parameters
$S Stop Stops processes
$Q Query Used to request all information from the current
node in the tree forwards up to and including the
values
$Q.P Path Used to request the path from the root of the tree
up to the current node
$Q.H Highest
Index
Used to request the number of daughter nodes of
the current node
$Q.N"i" Name Used to request the name of the daughter node
with index i, i = 1...n
$D Detail-Info Used to request detailed status information
$U qUit Used to terminate the data flow of the instrument,
e.g. after $Q
The triggers '$G' and '$S' are linked to particular objects, see Over-
view table in section 5.2.6.
All other triggers can always be used at all locations in the data tree.
Examples:
Request of the baud rate value: &Config.RSset.Baud $Q
Request of all values of the node 'RSset': &Config.RSset $Q
Request of the path of the node 'RSset': &Config.RSset $Q.P
Start mode: &Mode $G
Request of the detailed status: $D