Owner's manual
3.2 Operation via «IC Net»
762 IC Interface 23
Second press on RUN/STOP means 'stop' when measuring
If this option is enabled, the running data ac-
quisition will be stopped if key 33 or key 55
[ RUN/STOP ] on the 762 IC Interface front
panel is pressed. This is equal to Stop determi-
nation.
If this option is disabled, the running data ac-
quisition will not be stopped if key 33 or key 55
[ RUN/STOP ] on the 762 IC Interface front
panel is pressed.
<ASK> Read current parameters from 762 IC Interface.
<SET> Send current parameters to 762 IC Interface.
RS settings
The RS settings tab of the 762 IC Interface window contains RS232 pa-
rameters for the 762 IC Interface.
Baud Rate Baud rate of the device.
Selection: 1200...115200
Default value: 115200
Parity Parity check.
Selection: None, Even, Odd
Default value: None
Handshake Enable/disable software handshake mode.
Selection: Swchar, none
Default value: SWchar