3 Operation
762 IC Interface22
3.2 Operation via «IC Net»
This section describes only the most important points concerning the
operation of the 762 IC Interface. For further details please refer to the
«IC Net» Instructions for Use and to the on-line help in the PC pro-
gram.
3.2.1 Settings in the "762 IC Interface" window
By clicking the 762 icon in the toolbar with the left mouse button or by
clicking this icon with the right mouse button and selecting the Open
menu item the 762 IC Interface window for parameter settings is opened.
It consists of three tabs Init, RS settings, and Links.
Init
The Init tab of the 762 IC Interface window contains data acquisition pa-
rameters for the 762 IC Interface.
Label Optional label to name the interface with
maximum 8 characters.
ROM version Instrument program version number.
Measuring per second Number of data points measured per s.
Entry range: 10, 20, 30, 50, 60 points/s
Range Range for AD converter .
Example: converts ± 2500 mV to 2
24
bits
Entry range: 2500, 1250, 625, 312.5, 156.25, 78.125,
39.062 mV