Manual

5.1 Operation with manual sample change
761 SD Compact IC / Instructions for Use 8.761.1043
81
2 Allow shutdown.smt to run
Insert the free end of the sample aspirating tubing 3 into a
vessel filled with water.
Start the system by selecting
SYSTEM / Control / Start determi-
nation
(see Software Instructions for Use «IC Net 2.3», Section
4.3.3).
Let the system run to the end (duration: 7.10 minutes). High-
pressure pump and peristaltic pump are switched off auto-
matically at the end.
5.1.2 Control with «IC Cap» (manual operation)
Launch the «IC Cap» software and log in (see Section 2.12.4).
Configuration
manual.cfg should have been installed during installation
(see Section 2.12.2). If not, configuration
manual.cfg should now be
loaded (see Section 4.2.1).
5.1.2.1 Conditioning (manual operation / «IC Cap»)
The instrument should be conditioned after each restart:
1 Settings
Click on the yellow <Sample> button at the top right so that
it changes to the green <Start> button.
Choose System
startup.smt for item System.
Choose type "Sample" for item
Run Type.
Enter "Startup" for item
Ident.
2 Condition the instrument for 30 minutes
Insert the free end of the sample aspirating tubing 3 into a
container filled with water.
Start the system by clicking on the green <Start> button
and allow it run through to the end (30 minutes). Clicking on
the <Chrom> button displays the
CHROMATOGRAM win-
dow.
The baseline should be stable at the end.