Manual

5 Operation
761 SD Compact IC / Instructions for Use 8.761.1043
78
2 Condition the instrument for 30 minutes
Insert the free end of the sample aspirating tubing 3 into a
container filled with ultra-pure water.
Start the system by selecting
SYSTEM / Control / Start determi-
nation
(see Software Instructions for Use «IC Net 2.3», Sections
4.3.3 and 4.3.4) and allow it run through to the end (30 mi-
nutes).
The baseline should now be stable.
5.1.1.2 Calibration and sample determination (manual operation / «IC
Net»)
The instrument should be calibrated after each restart.
The sample aspirating tubing
3
should be immersed in the PE bottle
(6.1608.080) filled with ultra-pure water after each sample (or after an
MCAL standard).
1 Switch to system manual.smt and link it
Switch (see Software Instructions for Use «IC Net 2.3», Section
4.2) to system
manual.smt (see Section 3.2.2).