3.2 Systems supplied
761 SD Compact IC / Instructions for Use 8.761.1043
61
Links tab
Always connect the 766 IC Sample Processor to COM2.
Linked method
Method
SD_phosphate.mtw is linked to system auto.smt.
Most important settings:
• Always ensure that the path for
Chromatogram directory on the
Processing tab of the
METHOD SETUP window exists. Other-
wise, it must be adapted (see Section 3.1.3). The default path is
C:\Metrohm\SD Analyzer..
• Fields
IC Cap and Excel should be activated on the Export tab of
the
METHOD SETUP window.
• "4
min" should be entered for Duration on the General tab of the
METHOD SETUP window.
3.2.4 System "shutdown.smt"
System shutdown.smt is used after the actual measurements (at the end
of the working day) in order to shut down the instrument. Water is in-
jected and the runtime of the program is 7 minutes.
Always inject water as the last sample and thus rinse the tubing con-
nections with water in order to avoid damage to the instrument.