Manual
3.2 Systems supplied
761 SD Compact IC / Instructions for Use 8.761.1043
55
Configuration tab
Same settings as in the example with system
manual.smt in Section
3.1.6.
Links tab
Same settings as in the example with system
manual.smt in Section
3.1.6. The 761 SD Compact IC should always be attached to COM1.
Linked method
Method
SD_startup.mtw is linked to system startup.smt.
Most important settings:
• You should always ensure that the path for
Chromatogram directory
exists on the Processing tab of the
METHOD SETUP window.
Otherwise, it must be adapted (see Section 3.1.3). The default
path is C:\Metrohm\SD Analyzer..
• Field
IC Cap should be activated on the Export tab of the
METHOD SETUP window.
• "30
min" should be entered for Duration on the General tab of the
METHOD SETUP window.
3.2.2 System "manual.smt"
System manual.smt is used to manually measure samples (see also
Section 5.1).
System
manual.smt was used in Section 3.1 as an example for the
screenshots.
System window
See Section 3.1.2.
System parameters
See Section 3.1.6.
Linked method
Method
SD_phosphate.mtw is linked to system manual.smt.
Most important settings:
• You should ensure that the path for
Chromatogram directory on the
Processing tab of the
METHOD SETUP window exists. Other-
wise, it must be adapted (see Section 3.1.3). The default path is
C:\Metrohm\SD Analyzer..
• Fields
IC Cap and Excel should be activated on the Export tab of
the
METHOD SETUP window.
• "4
min" should be entered for Duration on the General tab of the
METHOD SETUP window.