3.1 «IC Net» – User interface for the 761 SD Compact IC
761 SD Compact IC / Instructions for Use 8.761.1043
53
PowerOn values Display of the values for the remote out-
put lines for switching on the instrument,
set on page
Hardware.
Pulse length Pulse length in ms.
Alarm stops
The
Alarm stops tab of the hardware settings window defines the events
for which the 761 SD Compact IC is stopped immediately. At an alarm
stop, high-pressure pump and peristaltic pump are stopped immedi-
ately, the running determination and the active sample queue are also
stopped.
Event Events for alarm stop:
Pump stop Pump stopped because pressure limits are ex-
ceeded.
Hardware error Hardware error detected at the 761 SD Com-
pact IC (high-pressure pump, injection valve, or
suppressor not working correctly).
Leak Leak detector has detected solvent in the in-
struments interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without connection
to the PC.
Power on Temporary power failure at the 761 SD Com-
pact IC hardware.