Manual
3 «IC Net»
761 SD Compact IC / Instructions for Use 8.761.1043
52
lution with a known conductivity through
the IC system. Observe the displayed
conductivity and change the cell constant
until the correct conductivity is displayed.
<Brake times> Option for changing the brake times for
the injection valve Valve and the suppres-
sor module
Suppressor.
Please change these values only after
consulting Metrohm Service.
Outputs
Automatic output of remote output signals for specific events is defined
on the
Outputs tab.
Event Events for automatic remote signal output:
Inject Switch over the injection valve to position "IN-
JECT".
Fill Switch over the injection valve to position "FILL".
Suppressor step Advance of the suppressor module to the next
position.
Pump stop (P max/min)
The pump stops owing to violation of the pres-
sure limits.
Hardware error Hardware error on 761 SD Compact IC (high-
pressure pump, injection valve or suppressor
not operating correctly).
Alarm leak detector Leak in the inner compartment.
Remote line Sets the output lines 1…8.
Selection: 0 (line off, inactive, open)
1 (line on, active, 0 V)
p (output of a pulse)
* (do not change state)