Manual

2.14 Rinsing before fitting the column
761 SD Compact IC / Instructions for Use 8.761.1043
37
2.14 Rinsing before fitting the column
Before the column is fitted, you should rinse again for 10 minutes. Dur-
ing this rinsing time, the connections should be checked for leaks and
the pressure of the peristaltic pump should be set optimally.
1 Check the fluid connections
Immerse the aspirating tubing for sample 3 into a sample
vessel filled with ultra-pure water.
Check whether the "tubing connections to MPaks" are con-
nected to the relevant MPak (eluent or regenerant).
Check that all waste outlet capillaries lead into a waste con-
tainer.
2 Open and connect the system
Launch the «IC Net» PC program if it has not already been
launched and open the
startup.smt system with File / Open /
System
unless it has already been opened.
Y
ou should see
the following window on the screen:
3 Connect the workplace
Choose item Connect to workplace from the Control menu in
this window.
4 Start the pump
Choose item Startup hardware from the Control menu in the
above window. All saved settings are restored.
5 Set the pressure for pump tubing
Push the pressure lever 58 up on both tubing cartridges 57
until regeneration and rinsing solution is aspirated.
Then push the pressure lever 58 one detent position up
w
ards
to achieve an optimum pressure.