1 Detach the column connection capillary 33 from the
PEEK coupling
Undo the pressure screw 50 with which the column connec-
tion capillary 33 is screwed to the PEEK coupling 54 (transi-
tion to the suppressor). Take this end of the capillary and
route it into a beaker.
The column connection capillary 33 should now connect the in-
jection valve to the beaker.
2 Open and connect the system
Open the startup.smt system, connect it and make the same
settings as described in Section 2.13.1.
3 Rinse the pulsation dampener
Click on the <On> button for IC pump to switch on the high-
pressure pump and rinse the pulsation dampener 30 filled
with isopropanol for approx. 10 minutes with eluent.
Click on the
<Off> button for IC pump to switch the high-
pressure pump back off again.
4 Connect the column connection capillary 33 back to
the PEEK coupling
Take the end of the column connection capillary 33 out of the
beaker and screw it back onto the PEEK coupling 54 (as a
transition to the suppressor) with a pressure screw 50.