2 Installation
761 SD Compact IC / Instructions for Use 8.761.1043
32
Enter cell constant
Now enter the cell constant (see Section 2.4) printed on the detector
block. Proceed as follows:
1 Open and connect a system
• Click on File / Open / System in the main window. In the window
which now opens, select file
startup.smt and click on <Open>.
You will see the following system window:
• Select the
Connect to workplace menu item of the Control menu
in this window.
2 Open the window for hardware settings
• Click with the right mouse button on the 761 image in the sys-
tem window and choose menu item
Hardware. You will see the
following window for the hardware settings:
3 Enter the cell constant
• Enter the cell constant printed on detector block 1.732.0420
in field
Cell constant.
• Click on
<OK> to save the settings and close the window.