Manual

2.10 Fitting the rear panel
761 SD Compact IC / Instructions for Use 8.761.1043
27
2.10 Fitting the rear panel
Now refit the rear panel which was removed during installation (Section
2.4). Place the incoming and outgoing cables and tubing at the rear
into the openings in the rear panel.
Figure 12: Rear panel
8 Opening for outlet capillaries
for discharge of eluent, regeneration
and sample solution
9 Opening for inlet capillaries
for supply of eluent, regeneration and
sample solution
10 Opening for detector cable
11 Knurled screws
for fastening the rear panel
Metrohm recommends the following distribution over the openings:
Connection Opening leads to
Suppressor outlet capillary for eluent ("Waste") 44
Suppressor outlet capillary for regenerant ("Waste") 46
Connection capillary 48
8
Waste container
Tubing connection to eluent MPak 23
Tubing connection to regenerant MPak 23
9
Eluent MPak
Regenerant MPak
Detector connection cable
10
Input
"Detector Block"
Then fit the rear panel, fit the upper 2 knurled screws 11 and firmly
screw all 4 knurled screws 11 tight.
Waste B
Waste A
11
9
10
8