Manual

2 Installation
761 SD Compact IC / Instructions for Use 8.761.1043
20
2.6.7 Connection suppressor waste
Route the "suppressor outlet capillary for eluent ("Waste")" 44 at the rear
out of the 761 SD Compact IC into a waste container.
2.7 Installing the regenerant path
The regenerant (100 mmol/L H
2
SO
4
) is aspirated with the peristaltic
pump from the regenerant MPak and forced into the suppressor via the
filter unit PEEK. From the suppressor, it is then routed into a waste con-
tainer (see Figure 9).
2.7.1 Fitting the pump tubing for regenerant
One channel of the two-channel peristaltic pump is used to pump the
regenerant. Fit a pump tubing 61 6.1826.110 as follows:
1 Disengage the tubing cartridges
Detach and disengage the two tubing cartridges 57 above
the pump drive 49 by pressing in the snap-action lever 59 on
the retaining bracket.
Press the pressure lever 58 down fully.
2 Fit the couplings 60
Fit one coupling 60 6.2744.110 onto each end of the pump
tubing 61 6.1826.110. Moistening the tip of the coupling
lightly will make it easier to slip it on.
3 Fit the pump tubing
Insert the fitted pump tubing 61 as shown in Figure 8 into the
tubing cartridge. The left-hand stopper 62 should latch into
the corresponding fixture on the left-hand side of the tubing
cartridge.
4 Reengage the tubing cartridge
Reengage the tubing cartridge with pump tubing 61 in the re-
taining bracket. Engage it at the left first and then press the
right-hand side down until the snap-action lever 59 engages.
Ensure that the pump tubing is not kinked when doing this.
Leave the other tubing cartridge outside. It is fitted with the
pump tubing for the sample path (see Section 2.8.1).