Manual

6.3 Faults and malfunctions
761 SD Compact IC / Instructions for Use 8.761.1043
105
Malfunction Cause Rectification
No feed of re-
generation or
rinsing solution
for the suppres-
sor
Contact pressure too
low
Leakage in the system
Defective pump tubing
Contamination of the
filter in the filter unit
PEEK 6.2821.120
Counter-pressure at
suppressor module too
high
Adjust contact pressure
(see Section 2.14)
Check connections
Replace pump tubing (see
Section 6.2.9)
Replace the filter
6.2821.130 (see Section
2.3.3)
Clean or replace the sup-
pressor (see Sections
6.2.6
6.2.8)