6.3 Faults and malfunctions
6.3.1 Error messages
If any type of malfunction occurs during operation of the 761 SD Com-
pact IC, this is shown by error messages in the PC program, which ap-
pear either in an error window or in the
SYSTEM STATE window.
Follow the instructions listed in the Error window and close this win-
dow with
<OK>.
You fill find further details of the error messages of the
SYSTEM STATE
window, their possible causes and the procedure for rectifying them in
the Software Instructions for Use «IC Net 2.3» Section 4.5.
6.3.2 Malfunctions and their rectification
If difficulties appear with The 761 SD Compact IC during analyses, their
causes are best investigated in The order separating column high-
pressure pump eluent connections. Several of The malfunc-
tions which may appear are listed in The following table with details of
possible causes and countermeasures.
Malfunction Cause Rectification
Baseline with high
noise level, pulsa-
tion
Contaminated pump
valves
Defective piston seals
Clean the valves (see Sec-
tion 6.2.5)
Replace the piston seals
(see Section 6.2.5)
Drift of the
baseline
Thermal equilibrium not
yet reached
Leakage in the system
Condition system with
heating switched on
Check connections and
seal them
Considerable
pressure drop
Leakage in the system Check connections and
seal them
Considerable
pressure rise
Blockage of filter unit
PEEK 6.2821.120
Change of column pack-
ing by injection of con-
taminated samples
Replace the filter
6.2821.130 (see Section
2.3.3)
Regenerate the column or
replace the column (see
Section 6.2.4)
Note
:
Samples should always
be microfiltered
Chromatograms
with poor resolu-
tion, Change in
the retention times
Deterioration in separa-
tion efficiency of the IC
column
Regenerate the column or
replace the column (see
Section 6.2.4)
Extreme peak
broadening,
splitting (double
peaks)
Dead volume at the col-
umn ends
Regenerate the column
(see Section 6.2.4)