Owner manual

4 Operation
761 Compact IC
86
Flow, mL/min Startup value for flow rate of the high-pressure
pump.
Entry range:
0.20 ... 2.50 mL/min
Pressure max, MPa Startup value for maximum pressure limit for
high-pressure pump.
Entry range:
0.0 ... 25.0 MPa
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump.
Entry range:
0.0 ... 25.0 MPa
Full scale, µS/cm
Startup value for full scale range.
Selection:
50, 250, 1000 µS/cm
Remote lines Startup value for remote line settings 18.
Selection:
0, 1
Instrument control for connected system
A double-click on the instrument icon or selection of the
Open menu
item using the right mouse button opens the system settings window.
For a connected system, the
Control tab is displayed. It allows manual
control of the 761 Compact IC functions and setting of startup values to
be sent to the instrument. This tab shows the current measurement val-
ues for conductivity and pressure.
Conductivity, µS/cm Live display of measured conductivity.
Pressure, MPa Live display of measured pressure.
The color settings for this two fields can be
changed by clicking the fields with the right