Owner manual
3 Operating tutorial
761 Compact IC
54
3.2 Preparations
Before you start this brief tutorial, the entire IC system must be correctly
installed as described in section 2. In what follows, the most important
points for the installation are described once again (for details, see the
sections mentioned).
1 Install 761 Compact IC
⇒ Setting up instrument section 2.2
⇒ Installing and connecting detector block section 2.3.1
⇒ Mounting syringe and aspirating tubing section 2.3.2
⇒ Mounting drain tubes section 2.3.3/4
⇒ Mains connection section 2.4
⇒ Connection to PC section 2.5
2 Prepare eluent
⇒ Preparing eluent:
2 mmol/L NaHCO
3
/ 1.3 mmol/L Na
2
CO
3
in dist. H
2
O
⇒ Microfiltering and degassing eluent section 5.1.3
3 Install high-pressure pump
⇒ Removing transport security screws section 2.6.1
⇒ Mounting pulsation dampener section 2.6.2
⇒ Installing eluent supply section 2.6.3
⇒ Degassing pump section 2.6.4
4 Connecting separating column and suppressor
⇒ Connecting IC anion column section 2.7.8
⇒ Connecting peristaltic pump section 2.8.2
⇒ Connecting supply bottles section 2.8.3
⇒ Connecting suppressor module section 2.8.4
⇒ Conditioning system section 2.9.2