Owner manual

2 Installation
761 Compact IC
28
3 Mount CO
2
absorber tube
First place a piece of cotton wool 70 followed by CO
2
ab-
sorber 69 (e.g. Merck soda-lime pellets with indicator, no.
6839.1000) in the large opening of absorber tube 72 and then
close it with the plastic lid.
Fasten absorber tube 72 to bottle attachment 66 with the aid
of SGJ clip 71.
4 Connect aspirating tubing to pump
Insert the free end of aspirating tubing 63 into one of the
openings 12 in the interior of the 761 Compact IC (see Fig. 2).
Pull aspirating tubing 63 sufficiently far into the interior of the
761 Compact IC, cut off to the required length and push at
least 5 mm of it onto aspirating capillary 40 (see Fig. 3 and
Fig. 4) of the high-pressure pump (it may be necessary to use
emery paper).
If necessary, fix aspirating tubing 63 in the required position
in the interior with the aid of the Y.107.0150 self-adhesive
strap.
Instead of threaded stopper
65
, the 4.420.0311 Tubing nipple (M6)
which belongs to the 6.1602.160 Bottle attachment accessories can
be used together with the included second E.301.0021 O-ring for
recycling operation (see section 5.2.3), in which the outlet capillary
from the detector block is led back into eluent bottle
67
.
63
64
65
66
72
68
71
70
69
67
Fig. 11
: Connection of eluent bottle
63 Aspirating tubing (6.1834.010)
64 Tubing nipple (4.420.4300; M8)
with E.301.0021 O-ring
65 Threaded stopper (6.1446.040; M6)
66 Bottle attachment (6.1602.105)
67 Eluent bottle (6.1608.070)
68 Aspirating filter (6.2821.090)
69 CO
2
absorber
70 Cotton wool
71 SGJ clip (6.2023.020)
72 Absorber tube (6.1609.000)