Select the Connect to workplace menu item of the Control menu
in this window.
If the connection between PC and Compact IC is working the
message
Hardware settings file for 761 unit with serial number
'#####' not found! Create?
will appear. Click <Yes> in order to
create the configuration file
#####.761 for this instrument.
If the connection between PC and Compact IC does not work
then the message
Detection of hardware failed[761 Compact IC
[ COM# ]]
appears in the SYSTEM STATE window. In this case
check whether the instrument has been switched on, whether
the connection cable is connected up properly and whether
the COM interface has been set correctly (see point 2). Then
repeat point 3.
4 Hardware settings
Of the general hardware settings only the input of the cell constant
is described. Standard settings can normally be used for all other
parameters.
Click the 761 icon using the right mouse button and select
the
Hardware item. The Hardware settings window is opened:
In the
Cell constant field enter the cell constant which is printed
on the 1.732.0110 Detector block (see section 2.3.1).
Click on
<OK> to close the window and save the settings.