5.3 Faults and malfunctions
761 Compact IC
187
Malfunction Cause Rectification
Chromatograms
with poor resolu-
tion, change in the
retention times
Deterioration in
separation efficiency of
the IC column
Regenerate column (see
section 5.1.1) or replace
column
Extreme peak
broadening,
splitting (double
peaks)
Dead volume at the
column ends
Fill column with glass
beads ( 100 µm) or
replace column
No feed of
regeneration or
rinsing solution
for suppressor
Contact pressure too
low
Leak in system
Defective pump tubing
Contamination of the
filter in the 6.2821.120
Filter unit PEEK
Counterpressure at
suppressor module too
high
Adjust contact pressure
(see section 2.9.2)
Check connections and
make leakproof
Replace pump tubing (see
section 5.2.9)
Replace the 6.2821.130
Filter (see section 2.3.6)
Clean or replace suppres-
sor (see section
5.2.6
5.2.8)