Owner manual

5.2 Maintenance and servicing
761 Compact IC
181
Set the flow at the high-pressure pump to 0.5 mL/min and
rinse the suppressor unit with 1 mol/L H
2
SO
4
for 5 to 10 min.
As the pressure falls, slowly increase the flow at the high-
pressure pump to 2 mL/min. Do not exceed a maximum
pressure of 2 MPa (20 bar).
Switch off high-pressure pump.
Switch suppressor to the next position using the
button.
Connect the inlet capillary 98 marked "H
2
O" to the inlet capil-
lary 26 using coupling 33 (see Fig. 10 and Fig. 16).
Set the flow at the high-pressure pump to 0.5 mL/min and
rinse the suppressor unit with 1 mol/L H
2
SO
4
for 5 to 10 min.
As the pressure falls, slowly increase the flow at the high-
pressure pump to 2 mL/min. Do not exceed a maximum
pressure of 2 MPa (20 bar).
Switch off high-pressure pump.
3 Connect suppressor to IC system
Connect inlet capillaries 98 and 99 to the peristaltic pump
(see section 2.8.4).
If the pressure problems persist, replace the suppressor rotor
(see section 5.2.9).
5.2.8 Cleaning the suppressor
It may be necessary to clean the suppressor in the following cases:
High counterpressure on the suppressor connection
tubing
Irremediable blockage of the suppressor (the suppres-
sor can no longer deliver solutions)
Irremediable obstruction of the suppressor (the sup-
pressor can no longer be switched to next position)
To clean the connection piece and the rotor, proceed as follows (see
Fig. 22):
1 Disconnect suppressor from IC system
Disconnect input capillary 96 of the suppressor module 47
from the separating column 81 (see Fig. 16).
Disconnect output capillary 97 from inlet capillary 45.
Disconnect inlet capillaries 98 and 99 from the filter units 96
(supply from peristaltic pump).