Owner manual
4.7 Batch reprocessing
761 Compact IC
167
4.7.3 Batch reprocessing queue editor
Open batch reprocessing queue editor window
REPROCESS / <Edit sample table>
This menu item opens the batch reprocessing editor program for edit-
ing the batch reprocessing table.
No Row number.
File name Name of the chromatogram file (read-only).
Method Name of the method file *.mtw used for re-
cording the chromatogram (read-only).
Ident User defined identifier for chromatogram. Will
be placed into appropriate passport field when
starting the chromatogram.
Vial Autosampler vial position to take sample from.
Will be placed into appropriate passport field.
Volume Injected volume of sample in µL. Will be placed
into appropriate passport field.
Dilution Sample dilution prior to injection. Will be placed
into appropriate passport field.
Amount Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
= C
c
• Amount
s
/ Amount
c
Will be placed into appropriate passport field.
Internal standard amount Concentration of the internal standard compo-
nent for relative concentration calculations. Will
be placed into appropriate passport field.
Calibration level Calibration level (see section 4.4.5) for the
sample: Level 0 stays for normal analysis, levels
1 and greater for calibration runs. Correct filling