4.6 Sample queue
761 Compact IC
157
4.6.2 Sample queue control
Sample queue overview table
761 COMPACT IC / File / Open / Sample queue
SYSTEM / System / Sample queue
These two menu items open the OPEN window for opening an existing
sample queue file
*.que or creation of a new sample queue file *.que by
entering a new name. After confirmation with
<OK> the sample queue
overview table is opened showing the current status of the sample
queue:
No Row number.
System System file to be used for the determination.
Ident User defined identifier for chromatogram. Will
be placed into appropriate passport field when
starting the chromatogram.
Vial Autosampler vial position to take sample from.
This value will not be transferred automatically
to the autosampler. Make sure that the vial
positions entered in this column are identical
to the vial positions used at the autosampler.
Inj-s Number of injections for the same vial position
(only 1 injection is permissible for Level ≥ 1).
This value will not be transferred automatically
to the autosampler. Make sure that the
numbers of injections entered in this column
are identical to the numbers of injections used
at the autosampler.