Owner manual

4.4 Methods
761 Compact IC
131
Comment User defined comment from the Comment tab of
the passport.
Method example for water determination.
GLP Log Method and Data GLP messages from the
Method Log and Data Log tabs of the passport.
METHOD GLP LOG
15/02/1999 17:35:47
Method created !
15/02/1999 17:35:56
Modified: Passport;
Method saved to disk file C:\IC761\Methods\s-03.mtw
DATA GLP LOG
Time constant: 0.100000
19/02/1999 11:26:38 Urs Waldburger
Modified: RAW DATA! Integration; Peaks;
Data saved to disk file C:\IC761\DATA\J2191106.CHW
23/02/1999 10:02:29 Urs Waldburger
Data saved to disk file c:\ic761\data\j2191106.CHW
More items to report
This part of the
Report options window contains a list of rarely used re-
port items that can be included into the report on the user's choice:
Acquisition Parameters used for data acquisition from the
from the
General, Measure and Filters tabs of the
Method setup window.
ACQUISITION PARAMETERS
Channels: 1
Method duration: 20.00min
Run duration: 20.01min
Measurements (method): 12006
Measurements (run): 12006
Freq.divisor: 1
Sampling: 10.00 pts/sec
Device: 761Compact IC
Interface:
Program before:
Program after:
Spikes filter: Yes
Median filter: No
slit: 0
Gauss filter: No
slit: 0
Integration Integration parameters and integration events.
INTEGRATION DEFAULTS
Channel: Cond
Delay: 2.60 min
Width: 5.00 sec
Broadening: 2.00
Slope: 5.00
Asymmetry: 1.50
MinArea: 0.00
MinHeight: 0.01
Rider ratio: 0.00