Owner manual
4 Operation
761 Compact IC
116
Force horizontal base back
Set horizontal baseline for a single peak. The
baseline is drawn backwards from the peak
end point. The intersection of the baseline with
the signal is defined as peak start point.
Cancel horizontal base back
Clear
Force horizontal base backmode.
Enable baseline Disable baseline
penetration penetration
Enable baseline penetration
Allow crossing of the signal by the baseline.
Disable baseline penetration
Clear
Enable baseline penetration mode. The
baseline is aligned to the bottom value of the
signal.
Integration events for integration parameters
Set width Sets new Width parameter that supersedes de-
fault linear growth of expected peak.
Set slope Sets new Slope value.
Set min height Sets new Minimum peak height value.
Set rider ratio Sets new value of Rider ratio parameter.