Owner manual

4.4 Methods
761 Compact IC
113
If an integration event is added or modified, the Edit integration event
window is opened where the following parameters can be set:
Time Start time for integration event in min.
Event List box with integration events to be selected
(see below).
Value Parameter value for selected integration event.
This field appears for events that demand input
of an additional parameter.
Integration events for peak detection
Disable detection Set this mode to stop detection of new peaks. If
a peak is available at the moment of the event it
is either finished (downslope peaks) or rejected
(upslope peaks).
Enable detection Clear Disable detection mode.
Enable negative peaks Enable detection of negative peaks. In some
cases this mode may result in instability of de-
tection algorithm.
Disable negative peaks Disable detection of negative peaks. This event
does not influence negative peaks which al-
ready started.
Disable peak reject Set this mode, if a small peak should not be re-
jected because of its flat apex.
Enable peak reject Clear Disable peak reject mode.
Integration events for peak start/end
Set peak start Force the beginning of a new peak. If a peak is
available at the moment of the event it is either
rejected (upslope) or terminated (downslope).
Set peak end Force peak end at the time of the event. Up-
slope peaks are rejected (except those born by
Set peak start
event), downslope peaks are ter-
minated.