Owner manual

4 Operation
761 Compact IC
112
<Suggest> This function sets up reasonable integration pa-
rameters in the following way (assumes that
several peaks are marked manually in the peak
editor mode (see section 4.5.4) and wrong
peaks deleted):
Slope is set to 3.
Asymmetry is set to 1.5.
Width and Broadening are calculated by fitting
a straight line through the measured peak
halfwidth values vs. retention time.
Width is
determined as the y value of this line at the
Delay time, Broadening is calculated as the y
value at the end of the chromatogram di-
vided by the y value at the
Delay time. If this
procedure does not provide adequate result,
Width is set equal to the minimum value of
peak width for the chromatogram,
Broadening is set to 1.
Events
Tab
Events of Integration parameters window for definition of integration
events. Integration events have a higher priority than integration pa-
rameters and are used for fine tuning of the integration process. They
should be used only in the case when problems cannot be solved by
tuning a set of parameters from the integration parameters window.
Number of events Number of integration events (read-only).
<Add> Add event to the list.
<Modify> Edit the selected event.
<Delete> Delete the selected event.
<Apply> Accept integration parameters and perform re-
integration.