4 Operation
761 Compact IC
100
Column
Tab
Column of passport with column information.
Number Serial number of column (max. 256 characters).
ID Internal diameter of the column in mm.
Length Length of the column in mm. This parameter is
used to calculate
Linear flow.
Packing material
Column Column description (max. 256 characters).
Particle size Particle size of the column in ยตm. This value is
used for the calculation of reduced theoretical
plate height
Reduced TP height.
Void volume Void volume for the column in %. Used to cal-
culate
Logarithmic index, Capacity factor and Lin-
ear flow
. It is calculated by the system in accor-
dance with the settings defined on
Method setup
/ Math
.
Precolumn
ID Internal diameter of the precolumn in mm.
Length Length of the precolumn in mm (set length to 0
if no precolumn is used).