3.1 Important Features
760 Sample Changer, Introduction
15
3 Introduction
3.1 Important Features
Overview
The 760 Sample Changer provides most of the important features
of the 730 Sample Changer:
• Interchangeable Metrohm standard sample racks
• Working, rinsing and shift heights can be specifically defined
• Automatic rack recognition by means of standard rack codes
• Optical beaker test
• Two stirrers (magnetic or rod stirrer) can be connected
• Operation with most Metrohm instruments via Remote I/O lines
LimitationsLimitations
• 4 fixed procedures that cannot be edited
• No RS 232 interface, no built-in pumps
• Simple keyboard with 3 keys and 2 LEDs (green and red)
• The turning speed of the rack cannot be varied
• No special rack types can be defined
Basic Settings
• Working, rinsing and shift heights can be defined for each rack
type
• Stirrer speed can be defined for both stirrers together
• One of 4 methods can be selected as the standard method