3.9 Connection of the 813 Compact Autosampler
757 VA Computrace – Hardware
43
3.9 Connection of the 813 Compact Autosampler
With the 813 Compact Autosampler connected to the 757 VA Computrace Stand,
max. 18 samples can be transferred to the measuring vessel at the 757 VA Com-
putrace Stand. After each measurement, the measuring vessel is rinsed by means
of two 772 Pump Units connected to a 731 Relay Box. For operation of this sample
changer and the automatic addition of standard addition and auxiliary solutions by
means of 765 Dosimats, the following instruments and accessory parts are needed
(see also
section 6.2
):
Quant. Order no. Instrument/Accessory
1 2.813.0020 813 Compact Autosampler for VA applications
1 2.731.0010 731 Relay Box
2 2.772.0010 772 Pump Unit
1 6.2141.150 Cable 813–757–731
1 6.5323.010 Rinsing equipment for VA
1…5 2.765.0010 765 Dosimat
1…5 6.3014.153 Exchange unit 5 mL (for addition solutions)
1…5 6.3014.213 Exchange unit 10 mL (for auxiliary solutions)
1 6.2141.080
Cable 757–2
×
765
or
1 6.9921.170
Cable 757–5
×
765
This section describes the procedure for the connection of 813 Compact Autosam-
pler, 731 Relay Box and 772 Pump Units. For the connection of 765 Dosimats, see
section 3.8
.
3.9.1 Electrical connection
Before any instruments are attached to the 757 VA Computrace Stand,
the 757 VA Computrace Stand must be switched off using the mains
switch 12.
The 813 Compact Autosampler is connected to the socket "Remote"
15
of the 757
VA Computrace Stand with the optionally available
6.2141.150 Cable
(see
Fig. 15
).
The second end of the 6.2141.150 cable is used to connect the 731 Relay Box. The
third end of the 6.2141.150 cable is used to connect 765 Dosimats with the
6.2141.080 cable (1...2 Dosimats) or the 6.9921.170 cable (1...5 Dosimats). At the
813 Compact Autosampler, "
Method 2
" must be set (procedure see
Instructions for
Use 813
).
At the connection "DC1" of the 731 Relay Box, a 772 Pump Unit is connected as
si-
phoning pump
, at the connection "DC2", a 772 Pump Unit is connected as
rinsing
pump
(see
Instruction for Use 731
and
772
). The output voltage of the 731 Relay
Box must be set to
+24 V
(default setting, see
section 2.3
,
Instruction for Use 713
).
Additionally, the following settings for remote address selection must be done:
DC1: 9
(Siphoning pump)
DC2: A
(Rinsing pump)