Owner manual

3 Installation
757 VA Computrace – Hardware
42
3.8.3 Changing the Exchange unit
The Exchange unit mounted on the 765 Dosimat can be changed only in the ex-
change position which is reached after filling. Please proceed as follows:
1 Fill exchange unit
At the start of the VA Computrace program, the exchange unit is automati-
cally filled. So this step is only necessary if the Dosimat has already been
used during the running program session.
Click on
or
MAIN WINDOW / Utility / Dosimat control
to open the
DOSI-
MAT CONTROL
window.
Select the desired Dosimat in the
Burette
field.
Click the
button to fill the exchange unit installed on the
Dosimat.
2 Change Exchange unit
Unscrew connection nipple of the PTFE tubing of the 4-way microtip
26
from connection
2
and take off old Exchange unit.
Mount new Exchange unit on Dosimat and screw connection nipple of the
PTFE tubing of the 4-way microtip
26
onto connection
2
of the flat stop-
cock on the Exchange unit.
3 Enter new volume of Exchange unit
If the volume of the new Exchange unit is different to the volume of the old
one, proceed as follows:
Select
MAIN WINDOW / Settings / General Settings.
Open the
Hardware
tab in
the
GENERAL SETTINGS
window.
Enter the new
Volume
of the Exchange unit installed on the 765 Dosimat.
Close the VA Computrace software by clicking on
or selecting
File /
Exit
.
Restart the VA Computrace software.
4 Initialize Dosimat(s)
Click on
or
MAIN WINDOW / Utility / Dosimat control
to open the
DOSI-
MAT CONTROL
window.
Select the desired Dosimat in the
Burette
field.
Click the
button to empty and refill the exchange unit in-
stalled on the Dosimat. Repeat this procedure two times.
Check if there are air bubbles left in the glass cylinder of the exchange
unit. If this is the case, repeat the flushing procedure by clicking the
button.
Close the
DOSIMAT CONTROL
window.