Owner manual

3 Installation
757 VA Computrace – Hardware
32
3.3.9 Cleaning the MME
If the mercury in the multi-mode electrode is contaminated and this leads to distur-
bances, the MME must be cleaned and refilled with ultrapure mercury. Proceed as
follows:
1 Remove multi-mode electrode from 757 VA Computrace
Stand
Unscrew FEP tubing
30
and
38
from the MME, disconnect electrode
cable
16
from MME.
Take multi-mode electrode
17
out of measuring head
20
while gently
tapping the MME to knock off any mercury drops on the glass capillary
into the measuring vessel.
2 Remove old mercury
Place multi-mode electrode
17
in a horizontal position in the electrode
holder
93
(see
Fig. 11-3
). The slotted screw
80
is now at the top.
Unscrew slotted screw
80
using a suitable coin.
Carefully turn MME and empty mercury through the threaded opening
79
into a waste container placed in the drip pan
92
. While doing so, gently
tap the glass capillary
88
and the MME to ensure that all mercury flows
out of the MME.
3 Dismantle MME
Unscrew retaining nut
87
.
Take glass capillary
88
out of opening
83
, the sealing ring
85
and the
locking ring
86
are removed at the same time. Remove these two parts
from the glass capillary
88
.
Remove insert ring
84
from the MME.
Unscrew slotted screw
75
with a suitable coin in an anticlockwise direc-
tion from screw thread
77
.
Screw locking ring
74
onto slotted screw
75
and pull out of the MME.
4 Dispose of old capillary
Connect filling tubing
94
with the tubing coupling
97
to the two gas wash
bottles
96
and the vacuum pump (see
Fig. 10
).
Insert glass capillary
88
(capillary end) in the filling cone
95
of the filling
tubing
94
.
Remove mercury from the capillary with the vacuum pump.
5 Clean MME
Clean inner compartments of the MME, contact pin
81
and the screw
threads
77
,
79
and
83
with a lint-free cloth.
Thoroughly rinse all inner compartments of the MME and the unscrewed
individual parts with dist. water and then dry with N
2
.
Do not use any organic solvents.