Owner manual

3.3 Multi-mode electrode (MME)
757 VA Computrace – Hardware
31
3.3.8 Changing the capillary
Contamination of the glass capillary can necessitate its replacement. In such a
case, proceed as follows:
1 Remove multi-mode electrode from 757 VA Computrace
Stand
Unscrew FEP tubing
30
and
38
from the MME, disconnect electrode
cable
16
from MME.
Take multi-mode electrode
17
out of measuring head
20
while gently
tapping the MME to knock off any mercury drops on the glass capillary
into the measuring vessel.
Place multi-mode electrode
17
in a horizontal position in the electrode
holder
93
(see
Fig. 11-3
).
2 Unscrew slotted screw 75
Using a suitable coin, unscrew slotted screw
75
out of screw thread
77
until the contact surface of the black O-ring at the Plexiglas wall (thin,
black stripe) is just visible below the metal thread
77
.
3 Dismantle old capillary
Position multi-mode electrode
17
vertically in the electrode holder
93
(see
Fig. 11-1
).
Undo retaining nut
87
completely by turning anticlockwise and lift up until
the lower part of the glass capillary
88
with the wide opening is visible.
Gently tap the glass capillary
88
to knock off any residual mercury in the
wide opening into the MME.
Press the retaining nut
87
downward with one hand and with your other
hand take glass capillary
88
completely out of the mount.
4 Dispose of old capillary
Connect filling tubing
94
with the tubing coupling
97
to the two gas wash
bottles
96
and the vacuum pump (see
Fig. 10
).
Insert glass capillary
88
(capillary end) in the filling cone
95
of the filling
tubing
94
.
Remove mercury from capillary with the vacuum pump.
5 Replenish mercury if necessary
Proceed as described in
section 3.3.2
.
6 Mount new capillary
Proceed as described in
section 3.3.3
.
7 Fill capillary
Proceed as described in
section 3.3.4
or
section 3.3.5
.