4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
76
.Cond: Conditioning ON/OFF
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
4.2.2.15. Mode.Parameter.Presel.IReq id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq value, unit, all, OFF
Mode.Parameter.Presel.ReqTitr ON, OFF
Automatic inquiries after the start of the determination. From such an inquiry,
the determination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 4.2.2.56) or with &M $G, see 4.2.2.1.
.ReqTitr: Current generation starts during requests (with ON).
4.2.2.16. Mode.Parameter.Presel.SampleUnit g, ...up to 5 ASCII
Method specific sample unit, i.e. when the method is loaded, the current unit
of the sample size is overwritten by the unit from the method.
4.2.2.17. Mode.Parameter.Presel.LimSmplSize.Status ON, OFF
Mode.Parameter.Presel.LimSmplSize.LoLim 0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
4.2.2.18. Mode.Parameter.Presel.Id1Text id1/C21, 10 ASCII characters
Mode.Parameter.Presel.Id2Text id2/C22, 10 ASCII characters
Mode.Parameter.Presel.Id3Text id3/C23, 10 ASCII characters
Text for sample identifications.
4.2.2.19. Mode.Parameter.Presel.Cell no diaph., diaphragm
Mode.Parameter.Presel.GenI 100, 200, 400, auto
.Cell: Type of generator electrode.
.GenI: Current at the generator electrode in mA. "auto" means that the
current is switched in the course of determinations.
Default: 400 mA for cells without diaphragm, auto for cell with
diaphragm.
4.2.2.20. Mode.Parameter.Presel.Oven COM1, COM2, no
If an Oven is connected, its results will be incorporated into the result report of
the Coulometer.
If there is no Oven connected via RS232, this parameter has to be on "no".
4.2.2.21. Mode.Parameter.Presel.ActPuls first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 132.
4.2.2.22. Mode.Def.Formulas.1.Formula H2O, CXX, RSX, +, -, *, /, (, )
Mode.Def.Formulas.1.TextRS 8 ASCII characters