User Manual

4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
70
&Assembly
Object Description Input range Reference
& Root
:
Assembly Assembly control
. GenEl Generator electrode
.Pulse Pulses $G 4.2.2.77.
.Length Length of pulses 0...2000 ditto
.Current Current of generator electrode 0, 100, 200, 400 ditto
. Meas Measuring of indicator electrode
.Status Status ON, OFF 4.2.2.78.
.Ipol Polarization current of electrode 2, 10, 20, 40 ditto
.Outputs I/O outputs
.AutoEOD Automatic output of EOD ON, OFF 4.2.2.79.
.SetLines Set I/O lines $G ditto
.L0 Signal on L0 active,inactive,pulse,OFF ditto
up to L13
.ResetLines Reset I/O lines $G ditto
.Stirrer Stirrer control
.Status Status ON, OFF 4.2.2.80.
. Bur Buret
.Empty Empties the buret $G,$S,$H,$C 4.2.2.81.
.Prep Prepares the buret $G,$S,$H,$C ditto
.Rates Rates
.Forward Forward rate
.Select Type of rate control digital 4.2.2.82.
.Digital Digital rate 0...150, max. ditto
.Reverse as for forward rate
.Select Type of rate control digital ditto
.Digital Digital rate 0...150, max. ditto
.Fill Fill $G,$H,$C 4.2.2.83.
.ModeDis Dispensing $G,$S,$H,$C 4.2.2.84.
.Select Type of dispensing control volume, time ditto
.V Volume to be dispensed 0.0001...0.1...9999 ditto
.Time Time to dispense 0.25...1...86 400 ditto
.VStop Limit volume 0.0001...9999, OFF ditto
.AutoFill Filling after each increment ON, OFF ditto