4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
89
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the end of the
determination can be switched off. Thus, for example, in conjunction
with a Coulometer several determinations can be performed in the same
beaker. Before AutoEOD is switched on, line 3 must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inactive"
means resetting of the signal, "pulse" means output of a pulse of app.
150 ms, "OFF" means the line is not operated, see also page 131.
Warnings:
If you have "AutoEOD" to "ON", an active line 3 is set to "inactive" by
the EOD pulse.
L6 is the line of the activate pulse. An active line 6 is set to "inactive"
by the activate pulse.
L5 is the error line. It is continuously controlled by the Coulometer pro-
gram and can therefore not be set freely.
Line assignments in Coulometer program:
L0 Ready, inactive state
L1 Conditioning OK
L2 Titration in progress
L3 EOD (End Of Determination)
L4 ---
L5 Error
L6 Activate pulse + can be set in TIP
L7 Pulses for recorder
L8 Connected remote box
L9,10 ---
L11 Change reagent
L12 Sample size out of limits
L13 Result out of limits
.ResetLines: Lines are set to the inactive status (= high).
4.2.2.80. Assembly.Stirrer.Status ON, OFF
Switching stirrer ON/OFF.
4.2.2.81. Assembly.Bur.Empty $G, $S, $H, $C
Assembly.Prep $G, $S, $H, $C
Starts the function "empty" and "preparation" resp. on the connected Dosino.
4.2.2.82. Assembly.Bur.Rates.Forward.Selected digital
Assembly.Bur.Rates.Forward.Digital 0...150, max.
Assembly.Bur.Rates.Reverse.Selected digital
Assembly.Bur.Rates.Reverse.Digital 0...150, max.
Expelling and aspirating rate in mL/min. "max." means maximum possible rate with
the Exchange Unit in current use.
4.2.2.83. Assembly.Bur.Fill $G, $H, $C