4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
87
Info.ActualInfo.Titrator.I read only
Info.ActualInfo.Titrator.Pot read only
Info.ActualInfo.Titrator.IPulse read only
$Q sends the current values.
.CyclNo: Cycle number of the voltage measurement cycle, e.g. "127". From the
cycle number and the cycle time (see 4.2.2.77), a time frame can be set
up. The cycle number is set to 0 on switching on the instrument and on
every start. It is incremented as long as the instrument remains switched
on.
.Water: Total water in ug, e.g. "62.313"
.Meas: Measured value in mV at the indicator electrode, e.g. "104.2".
.dWaterdt: Rate or drift in ug/min, e.g. "23.0".
.I: Total charge in mA·s, e.g. "667.48".
.Pot: Voltage at generator electrode.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
.IPulse: Current of actual pulse.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
OV will be sent for "overrange".
4.2.2.72. Info.ActualInfo.MeasPt.Index read only
Info.ActualInfo.MeasPt.X read only
Info.ActualInfo.MeasPt.Y read only
Info.ActualInfo.MeasPt.Z1 read only
Info.ActualInfo.MeasPt.Z2 read only
Info.ActualInfo.EP.Index read only
Info.ActualInfo.EP.X read only
Info.ActualInfo.EP.Y read only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry into
the list of EP.
.MeasPt.X Time in s, e.g. "14".
.MeasPt.Y Water in ug, e.g. "27.5".
.MeasPt.Z1 Measured value in mV, e.g. "160.3".
.MeasPt.Z2 Rate in ug/min, e.g. "100.5".
.EP.X Water in ug, e.g. "26.6".
.EP.Y Measured value in mV, e.g. "98.6".
4.2.2.73. Info.ActualInfo.Oven.HeatTime read only
Info.ActualInfo.Oven.SampleTemp read only
Info.ActualInfo.Oven.LowTemp read only
Info.ActualInfo.Oven.HighTemp read only
Info.ActualInfo.Oven.GasFlow read only
Info.ActualInfo.Oven.UnitFlow read only
$Q sends the current values from a connected KF Oven. If no Oven is connected, the
values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.