4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
71
&Setup
Object Description Input range Reference
& Root
:
Setup Settings for the operating mode
.Comport Output of automatic info 1,2,1&2 4.2.2.85.
.Keycode Send key code ON, OFF 4.2.2.86.
.Tree Sending format of path info
.Short Short format of path ON, OFF 4.2.2.87.
.ChangedOnly Paths of modified nodes only ON, OFF ditto
.Trace Message on changed values ON, OFF 4.2.2.88.
.Lock Lock key functions
.Keyboard Lock all keyboard keys ON, OFF 4.2.2.89.
.Config Lock <CONFIG> key ON, OFF ditto
.Parameter Lock <PARAM> key ON, OFF ditto
.SmplData Lock <SMPL DATA> key ON, OFF ditto
.UserMeth Lock functions
.Recall Lock “loading” ON, OFF ditto
.Store Lock “saving” ON, OFF ditto
.Delete Lock “deletion” ON, OFF ditto
.Exchange Lock <EXCH> key ON, OFF ditto
.Display Lock display function ON, OFF ditto
.Mode Setting waiting intervals
.StartWait Waiting time after start ON, OFF 4.2.2.90.
.FinWait Waiting time after run ON, OFF ditto
.SendMeas Automatic sending of measured values
.SendStatus Connect/disconnect sending ON, OFF 4.2.2.91.
.Interval Time interval 0.4...4...16200, ditto
MPList
.Select Selection Assembly, Titrator 4.2.2.92.
.Assembly From assembly
.CyclNo Cycle number ON, OFF 4.2.2.93.
.I Total charge (mA·s) ON, OFF ditto
.Meas Measured indicator voltage ON, OFF ditto
.Pot Voltage at generator electrode ON, OFF ditto
.IPulse I of current pulse ON, OFF ditto
.Bur Connected buret
.V Volume of dosing unit ON, OFF ditto
.Titrator From Titrator
.CyclNo Cycle number ON, OFF 4.2.2.94.
.Water Mass of water ON, OFF ditto
.Meas Measured indicator voltage ON, OFF ditto
.dWaterdt Drift or rate ON, OFF ditto
.I Total charge (mA·s) ON, OFF ditto
.Pot Voltage at generator electrode ON, OFF ditto
.IPulse I of current pulse ON, OFF ditto