User Manual

4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
65
.Set Setting of date and time $G 4.2.2.42.
.Date Date XXXX-XX-XX
.Time Time XX:XX
.RunNo Run number 0...9999 4.2.2.43.
.OpLevel Operator level standard, expert 4.2.2.44.
.StartDelay Start delay time 0...999 999 4.2.2.45.
.ResDisplay Result display standard,bold 4.2.2.46.
.DevName Device label 8 ASCII char. 4.2.2.47.
.Beep Beeper 1...3, OFF 4.2.2.48.
.DisplayMeas Display of measured values ON, OFF 4.2.2.49.
.Prog Program version read only 4.2.2.50.
.RSSet1 Settings RS232, 1 $G 4.2.2.51.
.Baud Baud rate 300,600,1200,2400,4800,
9600 ditto
.DataBit Number of data bits 7, 8 ditto
.StopBit Number of stop bits 1, 2 ditto
.Parity Parity even, odd, none ditto
.Handsh Handshake HWs, SWchar,
SWline, none ditto
.RSSet2 as for RS1
.Report Report configuration 4.2.2.52.
.Id Print report id ON, OFF ditto
.Instr Print line with instrument id ON, OFF ditto
.DateTime Print line with date/time ON, OFF ditto
.Run Print run number ON, OFF ditto
.Method Print line with method id ON, OFF ditto
.Sample Print line with sample size ON, OFF ditto
.Drift Print line with drift correction ON, OFF ditto
.TitrTime Print line with titration time ON, OFF ditto
.EPH2O Print line with H2O in ug ON, OFF ditto
.Statistics Print current statistics data ON, OFF ditto
.Visum Print line for visum ON, OFF ditto
.ComVar Values of common variables
.C30 C30 0... ±999 999 4.2.2.53.
up to C39 0... ±999 999