User Manual

4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
64
&Config
Object Description Input range Reference
& Root
:
Config Instrument configuration
: .Monitoring Monitoring functions
.Reagent Monitoring of reagent 4.2.2.31.
.Status Status of reagent monitoring ON, OFF ditto
.Determ Number of determinations 1...99...999, OFF ditto
.DCounter Determination counter 0...999 ditto
.MaxTime Time monitoring 1...7...9999, OFF ditto
.TCounter Time counter 0...9999 ditto
.ReagCap Reagent capacity monitoring 1...1000...9999, OFF ditto
.RCounter Reagent capacity counter 0...9999 ditto
.ClearCount Clears all counters above $G ditto
.Drift Change if drift is above 0...99, OFF ditto
.Change Change of reagent $G, $S 4.2.2.32.
.Status Type of reagent changing auto, man., OFF ditto
.WaitTime Waiting after switching off stirrer 0...999 999 ditto
.AspVol Volume to aspirate 0...100...9999 ditto
.SolventVol Solvent volume to add 0...100...9999 ditto
.Rinse Rinsing volume 0...9999 ditto
.NoRinse Number of rinsing cycles 1...9 ditto
.Validation Validation monitoring 4.2.2.33.
.Status Status of validation monitoring ON, OFF ditto
.Interval Time interval for validation 1...365...9999 ditto
.Counter Time counter 0...9999 ditto
.ClearCount Clears the counter above $G ditto
.Service Monitoring of Metrohm service $G 4.2.2.34.
.Status Status of service monitoring ON, OFF ditto
.Date Date of next service XXXX-XX-XX ditto
.DiagRep Printing of system test report ON, OFF 4.2.2.35.
.PeriphUnit Selection of peripheral units
.CharSet1 External printer on COM1 Epson,Seiko,Citizen
IBM,HP 4.2.2.36.
.CharSet2 as for CharSet1
.RepToComport Output of manual reports 1, 2, 1&2. And in 756:
int.,1&int.,2&int.,all 4.2.2.37.
.Balance Selection of balance Sartorius,Mettler,Mettler AT
AND,Precisa 4.2.2.38.
.Stirrer Stirrer control ON, OFF 4.2.2.39.
.RemoteBox Connected remote box 4.2.2.40.
.Status Status ON, OFF ditto
.Keyboard Type of keyboard US, deutsch, francais,
español, schweiz. ditto
.Barcode Input of barcode reader input, method, id1, id2,
id3, smpl size ditto
.Aux Miscellaneous
.Language Dialog language english, deutsch,
francais, espanol, italiano,
portugese, svenska 4.2.2.41.