User guide

4.2 Maintenance and servicing
754 Dialysis Unit
33
5
Close dialysis cell
Place top part 3838 on bottom part 3535 so that both parts lie
flush.
Screw the 5 screws 3939 in all the way in and tighten using the
6.2621.070 Allen key.
6
Attach dialysis cell
Place the assembled dialysis cell 3434 in cell holder 66 of the
754 Dialysis Unit.
7
Connect dialysis cell
Screw PEEK capillary 2424 to inlet opening 4242 of the bottom
part 3535 of the cell.
Screw PEEK capillary 2525 to outlet opening 4141 of the bottom
part 3535 of the cell.
Screw PTFE capillary 3333 to inlet opening 4343 of the top part 3838
of the cell.
Screw PTFE capillary 4545 to the outlet opening 4444 of the top
part 3838 of the cell.
8
Condition dialysis membrane
Immerse suction tubings 2626 and 2727 in the acceptor solution,
switch on pump with the <PUMP> key 1 1 and rinse the di-
alysis system with acceptor solution for approx. 20 min.
Check that the solution flows steadily from the two lines into
the waste bottle.
Check all tubings between the storage vessels, the tubing
cartridges, the dialysis cell and the waste bottles for leaks. If a
leak occurs anywhere, tighten or replace the connection in
question.
Switch off pump with the <PUMP> key 11.