User guide

4.1 Malfunctions and their rectification
754 Dialysis Unit
29
4 Malfunctions – Maintenance
4.1 Malfunctions and their rectification
If difficulties appear with the IC system during analysis, their cause are
best investigated in the order separating column pump eluent
IC system. An overview of possible malfunctions, their causes and
remedies can be found in the instruction manuals of the 732/733 IC
System (section 5.3.2). In addition to these general malfunctions the
table below contains others which might arise as a result of the opera-
tion of the 754 Dialysis Unit.
Malfunction Possible causes Rectification
Insufficient or non-
existing flow rate at
754 Dialysis Unit
Contact pressure too low
Pump tubing faulty
Inlets to dialysis cell blocked
Sample tubing blocked by
particles that are too large
Bacterial growth in the sample
channel and dialysis cell
Overly narrow PTFE tubings due
to excessive tightening of the
compression fittings
Set contact pressure correctly with contact
pressure lever (see section 2.3.3).
Replace pump tubing
(see section 4.2.5).
Replace the dialysis membrane and clean
the cell (see section 4.2.6).
Replace blocked tubing; always centrifuge
samples (see section 3.3.7).
Periodically replace tubings 2626, 3030, 3333 and
4545 for the sample channel; replace the
dialysis membrane
Replace tubings.
Leakage Leaking compression fittings Tighten the compression fittings or replace
if necessary.
Air bubbles in pump
circuit
Suction tubing is not immersed
Leaking compression fittings
Acceptor solution insufficiently
degassed
Standing air bubbles
Ensure suction tubing is fully immersed in
the acceptor or sample solution.
Tighten the compression fittings or replace
if necessary.
Re-degas the acceptor solution.
Small air bubbles are not a problem. If big
air bubbles occur, rinse the sample chan-
nel with degassed acceptor solution (open
outlet briefly if need be).
Insufficient yield from
dialysis
Transfer time incorrectly set
Dialysis time too short (e.g.
when concentration is too low)
Dried out or damaged dialysis
membrane
Deposits on the dialysis mem-
brane, recognizable from dis-
colouring at the sample inlet
Unsuitable dialysis membrane
Check optimum transfer time, especially
after each change of membrane or tubing
(see section 3.3.4).
Check optimum dialysis time (see section
3.3.5).
Replace dialysis membrane
(see section 4.2.6).
Replace dialysis membrane
(see section 4.2.6).
Use different dialysis membrane.