User guide

3.3 The dialysis procedure
754 Dialysis Unit
27
2
Measure the standard with dialysis
Set the dialysis time in the program to 5 min (see section
3.3.2).
Immerse sample suction tubing 2626 in the standard used in
point 1.
Start the program on the 732 IC Detector with the <PROG
R/S> key. Wait until the LED in the <PROG R/S> key stops
flashing, signalling the end of the evaluation.
3
Additional measurements
Increase the dialysis time in the program (see section 3.3.2)
by 5 min increments until the measured values remain con-
stant.
Take measurements as per point 2.
4
Determine optimum dialysis time
Record the ratio of the peak surfaces, i.e. concentrations
measured with and without dialysis in relation to the dialysis
time (i.e. the recovery rate) and determine the optimum dialy-
sis time.
3.3.6 Setting the program time
Set the program time so that the chromatogram can be completely re-
corded and evaluated within this period.
3.3.7 Dialysis procedure
To analyse samples using dialysis you should preferably proceed as
follows:
1
Prepare acceptor solution
To avoid malfunctions caused by air bubbles in the acceptor
channel, the ultrapure water used for the acceptor solution
should always be degassed for at least 10 min with vacuum,
N
2
or He.
2
Prepare sample
To prevent blockages in the sample channel, samples con-
taining a high volume of (suspended/solid particles) should
always be centrifuged using a table centrifuge at 10,000
rev./min for 5 min.