User guide

3 Operation
754 Dialysis Unit
22
3.2 Operation via 732 IC Detector
Switch instrument on/off
POWER
The 754 Dialysis Unit is switched on and off using mains switch
1313 on the rear of the instrument (see Fig. 2):
I Instrument switched on
0 Instrument switched off
After the instrument has been switched on the power lamp 33
lights up to show that the instrument is ready for use.
Switch on remote control of the pump
ON
OFF
REMOTE
In order to allow remote operation via 732 IC Detector, the remote
control must be switched on with switch 22.
ON Remote control switched on
OFF Remote control switched off
Switch pump on/off in a program
In the 732 IC Detector the <PROGRAM> key can be used to
create time programs with a maximum of 20 program steps (see
section 4.7.1 of 732/733 Manual). The pump can be switched on
or off at any program step by setting the remote output lead (lead
1 or 4 depending on the cable) to 1 (on, active, 0 V) or 0 (off,
inactive, open) respectively:
with 6.2143.200 cable with 6.2143.220 cable
Switch on:
>PROGRAM/edit XXX.X min
remote 1TTTTTTT
>PROGRAM/edit XXX.X min
remote TTT1TTTT
Switch off:
>PROGRAM/edit XXX.X min
remote 0TTTTTTT
>PROGRAM/edit XXX.X min
remote TTT0TTTT
Switch pump on/off at an event
In the 732 IC Detector the <EVENT> key can be used to
program a maximum of 4 different events (see section 4.7.3 of
732/733 Manual). The pump can be switched on or off at any
event by setting the remote output lead (lead 1 or 4 depending on
the cable) to 1 (on, active, 0 V) or 0 (off, inactive, open) respec-
tively:
with 6.2143.200 cable with 6.2143.220 cable
Switch on:
EVENT YY-MM-DD HH:MM:SS
remote 1TTTTTTT
EVENT YY-MM-DD HH:MM:SS
remote TTT1TTTT
Switch off:
EVENT YY-MM-DD HH:MM:SS
remote 0TTTTTTT
EVENT YY-MM-DD HH:MM:SS
remote TTT0TTTT