User guide
2.3 Attaching the accessories
754 Dialysis Unit
19
2.3.6 Attaching the accessories to the waste bottle
The waste bottle which receives the sample and acceptor solutions is
connected to the IC system as follows (see Fig. 8):
1
Attach bottle attachment
• Screw bottle attachment 4848 (6.1602.150) on to waste bottle
4949 (6.1608.070).
2
Connect tubing to the dialysis cell
• Pull one end of the PTFE capillary 45 45 connected to the dialy-
sis cell through a PVDF compression fitting 4646 (6.2744.000),
allowing 5 cm of tubing to project.
• Screw the PVDF compression fitting 4646 with tubing into an
opening in the bottle attachment 4848 and tighten so that the
tubing is held secure.
3
Connect tubing to valve A
• Pull the free end of the PEEK capillary 19 19 screwed to connec-
tion "2" of valve A through a PEEK compression fitting 4747
(6.2744.010), allowing 5 cm of tubing to project.
• Screw the PEEK compression fitting 4747 with tubing into an
opening in the bottle attachment 4848 and tighten so that the
tubing is held secure.
The end of the two tubings 1919 and 4545 inserted in the waste bottle
must be at the same height !